Scott Medical Products specializes in high accuracy, multi-component anesthetic gas mixtures used to calibrate and verify anesthesia monitors. Our anesthetic standards feature a guaranteed 2% relative blend tolerance and are available in a wide variety of refillable and nonrefillable cylinder sizes unmatched by any other medical gas manufacturer.
| Scott Medical Products offers agents including the following. Order toll-free 800-233-4334 or contact us. |
|
| Areas of Use |
- Operating room
- Intensive Care/NICU/CCU
- Biomedical engineering
- Instrument manufacturing
- Instrument service organizations
|
| SPECIFICATIONS |
Method of Preparation |
Concentration and Pressure |
Components |
Blend Tolerance |
Analytical Accuracy |
Gas Specifications |
Gravimetric |
≤ 1% Any Pressure |
Anesthetic CO2, O2, N2, N2O |
± 10% Relative |
± 10% Relative |
H2O < 20 ppm
CH4 < 25 ppm |
≥ 1%
40 ≥ psig |
Anesthetic
CO2, O2, N2, N2O |
± 2% Relative |
± 2% Relative |
H2O < 20 ppm
CH4 < 25 ppm |
≥ 1%
40 > P ≥ 10 psig |
Anesthetic
CO2, O2, N2, N2O |
± 5% Relative |
± 5% Relative |
H2O < 20 ppm
CH4 < 25 ppm |
≥ 1%
< 10 psig |
Anesthetic
CO2, O2, N2, N2O |
± 10% Relative |
± 10% Relative |
H2O < 20 ppm
CH4 < 25 ppm |
| Agent type |
Expiration Period |
Raw Material |
Traceability |
| Halocarbon-23 Mixtures |
BD Cylinders — 2 Years |
USP/NF or Equivalent |
NIST Class 1 Weights |
| Halothane Mixtures |
BD, H, K, TD, UD, XL Cylinders — 2 Years |
USP/NF or Equivalent |
NIST Class 1 Weights |
| SD Cylinders — 18 Months |
USP/NF or Equivalent |
NIST Class 1 Weights |
| ED Cylinders — 1 Year |
USP/NF or Equivalent |
NIST Class 1 Weights |
N2O and
Halocarbon-22 Mixtures |
AL, ED, H, K, PD, UD, XL Cylinders — 3 Years |
USP/NF or Equivalent |
NIST Class 1 Weights |
| BD, SD Cylinders — 2 Years |
USP/NF or Equivalent |
NIST Class 1 Weights |
| Sevoflurane Mixtures |
AL,BD, SD Cylinders — 2 Years |
USP/NF or Equivalent |
NIST Class 1 Weights |
| PD, TD, UD, XL Cylinders — 1 Year |
USP/NF or Equivalent |
NIST Class 1 Weights |
| All Other Anesthetic Mixtures |
AL, BD, H, K, SD, TD, UD, XL Cylinders — 2 Years |
USP/NF or Equivalent |
NIST Class 1 Weights |
| ED, PD, ZD Cylinders — 1 Year |
USP/NF or Equivalent |
NIST Class 1 Weights |
Anesthetic Pressure Information
In gas mixtures containing vaporized anesthetic agents such as halothane, the anesthetic agent can condense and separate back into a liquid if exposed
to a temperature below the mixture’s dew point. The table below lists the vapor pressures of the common anesthetic agents at 0°C and the maximum pressure for a mix of a given standard composition.
To calculate the restricted pressure of your mixture, divide the VP constant for the anesthetic by the concentration. For example, to calculate the
final pressure of a 1% halothane balance nitrogen gas mixture, you would divide 1.87 by 0.01 (1%). The resulting pressure or 187 psia (172 psig) is the
maximum fill pressure for your gas mixture. This product must be kept above 0°C or condensation and a resulting loss of concentration will occur.
| Agent |
Tradename |
VP @
0ºC |
Molecular
Weight |
Density
(g/ml) |
Pressure (psig)@ |
| 0.5% |
1.0% |
2.0% |
3.0% |
5.0% |
7.5% |
25% |
50% |
| Desflurane |
Suprane® |
4.93 |
168.04 |
1.47 |
971 |
478 |
231 |
149 |
83 |
51 |
– |
– |
|
Enflurane |
Ethrane® |
1.28 |
184.5 |
1.52 |
241 |
113 |
49 |
27 |
11 |
– |
– |
– |
| Halocarbon-22 |
Freon-22® |
72.2 |
86.47 |
– |
2000 |
2000 |
2000 |
2000 |
1429 |
947 |
273 |
129 |
| Halocarbon-23 |
Freon-23® |
363 |
70.014 |
– |
2000 |
2000 |
2000 |
2000 |
2000 |
2000 |
1452 |
726 |
| Halothane |
Fluothane® |
1.87 |
197.4 |
1.87 |
360 |
172 |
78 |
47 |
22 |
10 |
– |
– |
| Isoflurane |
Forane® |
1.77 |
184.5 |
1.50 |
339 |
162 |
73 |
44 |
20 |
8.9 |
– |
– |
| Nitrous Oxide |
– |
460 |
44.013 |
– |
2000 |
2000 |
2000 |
2000 |
2000 |
2000 |
1825 |
905 |
| Sevoflurane |
Ultane® |
1.08 |
200.06 |
1.52 |
201 |
93 |
39 |
21 |
6.9 |
– |
– |
– |
Order toll-free 800-233-4334 or contact us.
|